CHEMOSPHERE, vol.303, 2022 (SCI-Expanded)
For the quantifiable amounts of Telmisartan (TLM) and Hydrochlorothiazide (HYD) in the presence of Amlodipine (AML) in a ternary mixture of synthetic laboratory mixture, a novel, sensitive, quick, and practical reversed-phase high-performance liquid chromatography (RP-HPLC) method was given. In order to separate, a Waters Spherisorb ODS-2 C18 column was used. For HYD, TLM, and AML, these techniques were viable over linearity ranges of 4-12 mu g/mL, 4-25 mu g/mL, and 5-40 mu g/mL, respectively. The mobile phase system was acetonitrile:methanol: phosphate buffer at pH 2.5 (65:5:30 v/v/v), and the flow rate was 1.5 mL/min. Novel spectrophotometric methods were applied for active substances to determine simultaneously. The first method is absorptivity centering using factorized spectrum, and the second method is dual amplitude difference coupled